How Vickers Hardness Is Measured: Step-by-Step Guide

How Vickers Hardness Is Measured: Step-by-Step Guide

·

Vickers hardness testing remains one of the most accurate methods for measuring a material’s resistance to indentation. The technique is widely applied across metals, alloys, coatings, and thin samples where precision and surface sensitivity are essential.

The process involves applying a controlled load through a diamond pyramid indenter and then measuring the size of the resulting impression. The calculated value, known as the Vickers Hardness Number (VHN), provides a reliable indicator of material strength and surface quality — making it a preferred standard for both research and industrial quality control.

This article provides a complete explanation of how Vickers hardness is measured, covering the equipment setup, testing steps, calculation formula, and typical applications.

Understanding the Vickers Hardness Testing Principle

The Vickers hardness test method operates on a simple yet highly accurate principle: a diamond pyramid-shaped indenter is pressed into the material’s surface under a specified load, and the resulting indentation is measured using an optical microscope.

Key Parameters in the Vickers Method:

  • Indenter: Diamond pyramid with an angle of 136° between opposite faces.
  • Applied Load: Typically between 1 to 100 kgf (or as low as 10 g for microhardness testing).
  • Dwell Time: Usually 10–15 seconds to ensure stable penetration.
  • Measurement: The diagonal length of the indentation is measured optically.

The Vickers test is unique because it can be applied to all materials — from soft metals to hard ceramics — and uses one consistent indenter geometry regardless of load.

Advantages:

  • High precision and repeatability
  • Suitable for both macro and microhardness testing
  • Minimal sample damage
  • Conforms to ISO 6507 and ASTM E384 standards

By analyzing the indentation geometry, engineers can assess not only surface hardness but also the microstructural uniformity of the tested material.

Equipment and Setup: Diamond Indenter and Optical System

The accuracy of how Vickers hardness is measured depends largely on the quality of equipment and proper setup.

Main Components of a Vickers Hardness Tester:

  1. Diamond Pyramid Indenter – A four-sided, polished diamond tip forming a 136° angle.
  2. Load Application System – Applies a controlled load through mechanical, dead-weight, or electronic mechanisms.
  3. Optical Microscope – Used for diagonal length measurement of the indentation.
  4. Digital Display or Software Interface – Calculates and records Vickers Hardness Numbers (VHN)automatically.

Sample Preparation Requirements:

  • Smooth and polished surface (free from scratches or oxidation)
  • Clean and free of debris or oil residue
  • Stable positioning under the indenter

Calibration Considerations:

  • Regular calibration of the optical system and loading mechanism is essential.
  • Surface preparation directly influences light reflection and optical clarity, which affect indentation visibility.

Modern systems, such as those from Qualitest, feature digital imaging and automated indentation measurement, improving testing speed, accuracy, and consistency for both microhardness and macrohardness testing applications.

Step-by-Step Process: From Loading to Indentation Measurement

To ensure accurate and repeatable results, the Vickers hardness test procedure follows these sequential steps:

1. Surface Preparation

  • Polish the sample to achieve a mirror-like finish.
  • Ensure no contamination or irregularities on the surface.

2. Load Application

  • Select the appropriate applied load (typically 1–100 kgf, or lower for microhardness).
  • The diamond pyramid indenter is pressed vertically onto the specimen for a defined dwell time (10–15 seconds).

3. Indentation Formation

  • The applied load creates a square-shaped indentation.
  • Once the load is released, the indentation remains for optical measurement.

4. Optical Measurement

  • Use a calibrated microscope to measure the two diagonals of the indentation in millimeters (d₁ and d₂).
  • Calculate the mean diagonal length (d).

5. Hardness Value Calculation

  • Insert values into the Vickers formula:
    HV=1.854×Fd2HV=1.854×d2F​
    where F = applied load (kgf), and d = average diagonal (mm).

This formula calculates the Vickers Hardness Number (VHN) or Vickers Hardness Value (HV).

6. Record and Verify Results

  • Compare readings against known material hardness standards.
  • Repeat measurements at multiple points to ensure consistency.

How to Calculate the Vickers Hardness Number (VHN)

The calculation of Vickers Hardness Number (VHN) relies on precise measurement and accurate data input.

Formula:

HV=1.854×Fd2HV=1.854×d2F​

Where:

  • HV = Vickers Hardness Number
  • F = Applied test load in kilograms-force (kgf)
  • d = Mean diagonal length in millimeters (mm)

Example Calculation:
If a 10 kgf load is applied and the average diagonal length is 0.5 mm:

HV=1.854×10(0.5)2=74.16HV=1.854×(0.5)210​=74.16

Thus, the Vickers Hardness Value (HV) is approximately 74 HV10 (the subscript indicates the test load).

Measurement Tips:

  • Ensure the optical focus is sharp and uniform lighting is used.
  • Use digital imaging software to minimize human error in diagonal measurement.
  • Calibrate using ISO 6507 reference blocks for consistent accuracy.

The Vickers method’s ability to correlate with other hardness scales (Brinell, Rockwell) via a hardness conversion chart makes it ideal for cross-standard comparison in material testing laboratories.

Typical Applications and Materials for Vickers Hardness Testing

The Vickers test method is versatile and applicable to a wide range of materials, from metals and ceramics to coatings and composites.

Common Applications:

  • Microhardness Testing: Ideal for thin coatings, case-hardened surfaces, and microcomponents.
  • Metallurgical Research: Evaluates grain structure and microstructure uniformity.
  • Quality Control: Ensures materials meet hardness and durability specifications.
  • Heat Treatment Verification: Checks hardness variations after quenching, tempering, or annealing.
MaterialTypical Load RangeVickers Hardness (HV)
Aluminum0.5–1 kgf15–50 HV
Mild Steel10 kgf120–180 HV
Hardened Steel30 kgf400–700 HV
Tungsten Carbide50 kgf1500–2000 HV

Benefits of Vickers Testing:

  • Works on virtually all materials and thicknesses
  • Provides a continuous scale of hardness values
  • Non-destructive for micro-level evaluation

Because of its precision, the Vickers hardness test procedure is the preferred choice for microstructure evaluationin aerospace, automotive, and materials research industries.

Why Choose Qualitest for Accurate Vickers Hardness Testing Systems

For laboratories and manufacturers that require unmatched accuracy, Qualitest offers advanced Vickers hardness testing systems built for precision, reliability, and compliance with international standards.

Key Features of Qualitest Vickers Testers:

  • Automated load application and optical measurement for consistent results
  • Integrated digital imaging and indentation analysis tools
  • Full compliance with ISO 6507, ASTM E384, and other international standards
  • Compatible with microhardness and macrohardness testing configurations
  • Intuitive interfaces featuring data logging and automated reporting functions

Why Professionals Choose Qualitest:

  • Exceptional microscope accuracy and optical calibration stability
  • Consistent measurement repeatability across various sample types
  • Wide selection of models for research labs, QA facilities, and production floors
  • Supported by a global network for training, installation, and maintenance

From fine coating evaluations to macrohardness testing of heavy alloys, Qualitest delivers precision instruments engineered to produce reliable, repeatable, and standards-compliant results every time.
Contact Qualitest to learn more about our full range of Vickers hardness testing solutions.

Related Products