Coating Thickness Gauge QualiCTG™ I
Substrate mix-ups and over-limit spots can hide in routine coating checks. QualiCTG I automatically distinguishes base metal type, measures up to 2,000 µm with a ruby wear-resistant probe, and flags out-of-range results, with standards support on the page for ASTM E376 and ISO 2808.
Description
Applications
Features
Specifications
Frequent substrate switching in multi-material coating inspection causes setup errors and slows downstream production lines. The QualiCTG™ I addresses this challenge by combining magnetic induction and eddy current sensors into a compact, pocket-sized non-destructive testing instrument.
Engineered for technical precision, this gauge provides automatic substrate property recognition across ferrous and non-ferrous bases. It ensures strict compliance with ISO 2178, ISO 2360, and ASTM D7091 standards when measuring paints, galvanized layers, and thin plastic films.
Featuring a durable ruby wear-resistant probe tip and a manual rotation OLED display, the device records rapid 0.5-second inspection cycles. Its internal 320-measurement database exports directly via USB, maximizing throughput while delivering highly repeatable quality control reports.
The Coating Thickness Gauge QualiCTG™ I is utilized for rapid elemental verification and thickness inspection across various industrial and field environments:
- Automotive Inspection: Performing car paint testing to identify repaired panels or ensure paint thickness consistency.
- Anti-Rust Treatment Verification: Monitoring the effectiveness of galvanized layers and anti-corrosion treatments on structural steel.
- Hardware Parts Processing: Inspecting the coating thickness of machined components and functional hardware parts.
- Non-Conductive Cladding Measurement: Evaluating the thickness of claddings on ferromagnetic metals such as iron, nickel, and cobalt.
- Mixed Substrate Inspection: Measuring paint or plastic film thickness on aluminum and copper surfaces using eddy current detection.
- Film Thickness Detection: Verifying the uniformity of plastic films and industrial coatings during the manufacturing process.
- Ferromagnetic Metal Testing: Specifically measuring coating or plating thickness on iron and stainless steel surfaces.
The QualiCTG™ I combines high-speed detection with portable ergonomics to support efficient quality control:
- Integrated Dual-Principle Probe: Features a built-in magnetic induction sensor and an eddy current probe for comprehensive Fe and NFe testing.
- Automatic Substrate Recognition: Detects the properties of the base metal automatically, allowing for seamless transition between different materials.
- Enhanced Probe Durability: Equipped with a ruby wear-resistant probe to ensure long-term precision and minimize damage to sensitive coatings.
- Fast Reaction Performance: Provides immediate thickness readings with a reaction time of less than 0.5 seconds.
- Clear Statistical Visualization: Displays key metrics including the number of measurements, maximum, minimum, mean values, and standard variance.
- Configurable Measurement Range: Measures from 0 to 2000 μm with a high resolution of 0.1 μm for coatings under 100 μm.
- Intelligent Warning System: Features a red backlight over-limit alarm to alert operators when measurements exceed user-defined thresholds.
- Comprehensive Calibration Support: Supports both zero-point and multi-point calibration to maintain accuracy across varied surface conditions.
- Digital Data Management: Stores up to 320 measurement results and supports USB data transmission for PC-based reporting.
- Field-Ready Design: Includes a manual rotation screen for viewing at different angles, auto shut-off for battery protection, and an IP40-rated ABS housing.
Fe and NFe Coating Measurement Workflow
The QualiCTG™ I utilizes two distinct physical principles to measure applied layers without destroying the specimen. For Fe substrates—including ferromagnetic metals like iron, nickel, and cobalt—the gauge employs magnetic induction. For NFe substrates like aluminum or copper, it uses the eddy current effect. The integrated probe allows the device to distinguish between these substrate types automatically upon contact. To perform a check, the operator simply places the ruby probe against the surface; the thickness of the paint, galvanized layer, or plastic film is calculated instantly as the probe completes the circuit.
Data Display, Storage, and USB Workflow
Inspection documentation is supported by an intuitive dot array LCD and a streamlined menu interface. The system tracks statistical data automatically, allowing technical teams to review the variance and consistency of a coating batch in real-time. For permanent records, the QualiCTG™ I stores 320 individual measurements in its internal memory. These records can be transmitted to a computer via the included USB cable. Furthermore, the USB interface supports firmware upgrades, ensuring the instrument remains compatible with the latest analysis protocols and software refinements.
Calibration, Probe, and Measurement Conditions
Consistent precision is maintained through flexible user calibration options, including one-touch zero calibration and multi-point adjustment. The ruby probe is designed for a trigger force between 0.3 and 0.8 N, providing stable contact without marring the surface. To ensure reliable results, the instrument requires a minimum measurement area of 15 mm in diameter and adheres to specific substrate thickness limits: 0.30 mm for Fe and 0.05 mm for NFe. The unit operates reliably in temperatures ranging from -10 to +50°C, making it suitable for both factory floor and outdoor on-site inspections.
Measurement System
| Feature | Specification |
| Measuring Principle | Fe: Magnetic Induction / NFe: Eddy Current |
| Measuring Range | 0 to 2000 μm |
| Measurement Accuracy | ±(2.5% + 1 μm) |
| Resolution (0-99.9 μm) | 0.1 μm |
| Resolution (≥100 μm) | 1 μm |
| Measuring Units | μm, mm, mil |
| Reaction Time | < 0.5 seconds |
Calibration, Statistics, and Data
| Feature | Specification |
| User Calibration | Zero Calibration, Point Calibration |
| Statistical Values | Max, Min, Mean, Standard Variance, Count |
| Data Storage | 320 Measurements |
| Data Transmission | Supported via USB |
| Firmware Upgrade | Supported via USB |
Probe and Substrate Requirements
| Feature | Specification |
| Probe Type | Ruby Wear-Resistant |
| Probe Trigger Force | 0.3 to 0.8 N |
| Min. Radius (Convex) | 5 mm |
| Min. Radius (Concave) | 25 mm |
| Min. Measurement Area | Diameter 15 mm |
| Min. Substrate (Fe / NFe) | 0.30 mm / 0.05 mm |
Physical and Environmental
| Feature | Specification |
| Display | Dot array LCD with manual rotation |
| Operating Temp / Humidity | -10 to +50°C / 20 to 90% RH |
| Protection Level | IP40 |
| Housing Material | ABS |
| Battery Type | 2x 1.5V Alkaline or 1.2V Rechargeable |
| Weight / Dimensions | 80 g (without battery) / 114 × 53 × 25 mm |
Experience the Qualitest advantage and see how we can elevate your testing and quality control processes.
Please contact us for pricing, check availability, or ask for additional information about: