Home / Products / Other Portable Testers / Coating Thickness Gauge – Paint Thickness Gauge / Coating Thickness Gauge – QualiCTG™ V
Coating Thickness Gauge – QualiCTG™ V

Coating Thickness Gauge – QualiCTG™ V

Uneven probe pressure can distort coating data across broad surfaces. QualiCTG V uses gravity-based positioning for steadier contact, auto-detects ferrous or non-ferrous substrates, scans every 0.1 seconds, and feeds Bluetooth-linked records for faster paint, anodizing, and plating inspection.

Description

Applications

Features

Specifications

Manual substrate configuration errors and uneven probe pressure frequently cause distorted coating thickness data. The QualiCTG™ V Coating Thickness Gauge solves these paint, anodizing, and plating inspection bottlenecks with automatic Fe/NFe substrate detection and a unique gravity-stabilized contact design.

Featuring an ultra-durable tungsten carbide tip, this portable non-destructive testing (NDT) instrument ensures long-term calibration stability. Integrated Bluetooth syncs real-time data to mobile apps, while dual LCD screens allow seamless visibility from restricted, awkward, or inverted inspection angles.

Fully compliant with ISO 2178, ISO 2360, and ASTM D7091 standards, the gauge delivers traceably accurate readings within a 0.0 to 2000 μm range. It provides rapid 0.3-second evaluation for automotive, aerospace, PCB, and metalworking quality control workflows.

The Coating Thickness Gauge – QualiCTG™ V is utilized to verify coating integrity and finish quality across diverse manufacturing and processing sectors:

  • PCB Board Measurement: Inspecting the thickness of non-conductive layers and protective coatings on circuit boards.
  • Aluminum Profile Inspection: Verifying the consistency of anodized layers and paint coatings on architectural or industrial aluminum extrusions.
  • Car Paint Coating Measurement: Performing rapid automotive audits to identify paint thickness variations or repair work.
  • Metalworking and Manufacturing: Monitoring the thickness of varnish, enamel, chrome, and galvanized layers on steel and iron components.
  • Copper Foil Measurement: Evaluating the thickness of thin copper foils and coatings in electronics and battery manufacturing.
  • Chemical Processing Workflows: Validating anti-corrosion coatings on equipment used in aggressive environments.
  • General Quality Inspection: Supporting batch sampling and final product validation for hardware parts and finished consumer goods.
  • Substrate-Specific Evaluation: Measuring non-conductive coatings on non-magnetic metals including copper, aluminum, die-cast zinc, and brass.

The QualiCTG™ V combines robust hardware with intelligent digital tools to enhance inspection accuracy and efficiency:

  • Automatic Substrate Detection: Features an auto-sensing probe that identifies the base material (Fe or NFe) and selects the appropriate measurement principle without manual user input.
  • Gravity-Based Measurement Design: Utilizes the instrument’s own weight to apply consistent pressure, ensuring stable probe contact and reducing operator-induced variability.
  • Dual-Screen Visualization: Equipped with both front-facing and top-mounted displays, allowing operators to read results clearly from various perspectives during inspection.
  • Tungsten Carbide-Tipped Probe: Features a cemented carbide probe tip for exceptional wear resistance, maintaining calibration stability over millions of readings.
  • Bluetooth and Mobile App Integration: Supports real-time measurement transmission to a mobile device for advanced data analysis and report sharing.
  • Rapid Inspection Workflow: Delivers readings at a 0.3-second technical interval, with a specialized scanning mode capable of capturing data points every 0.1 seconds.
  • Sophisticated Calibration Control: Supports zero-point adjustment and comprehensive 1 to 5 point calibration using included zero plates and standard films.
  • Live Thickness Tracking: Provides real-time tracking of maximum and minimum thickness values during surface scanning to identify deviations immediately.
  • Onboard Data Storage: Stores up to 9 complete measurement sets directly on the unit for convenient field review.
  • Integrated Temperature Compensation: Maintains measurement precision in varying environmental conditions through automatic thermal adjustment.

Fe and NFe Coating Measurement Workflow

The QualiCTG™ V employs two distinct physical principles to measure applied layers non-destructively. For Fe substrates—such as steel and iron—the gauge utilizes magnetic induction to evaluate non-magnetic coatings like paint, varnish, and chrome. For NFe substrates—such as copper, aluminum, and brass—the instrument employs the eddy current effect to measure non-conductive coatings including anodized layers and ceramic finishes. The integrated probe identifies the base metal properties upon contact, allowing the technician to move seamlessly between different components while the gauge automatically updates its calculation model.

Calibration, Probe Contact, and Rapid Scanning

Operational repeatability is supported by the QualiCTG™ V’s gravity-based measurement system. By allowing the weight of the device to stabilize the contact point, the instrument minimizes errors caused by uneven manual pressure. For specialized inspection tasks, technical teams can utilize the 1 to 5 point calibration workflow to ensure accuracy across specific thickness ranges. During large-surface audits, the 0.3-second measuring interval supports rapid scanning; the gauge captures and displays the peaks and valleys of the coating thickness in real-time, helping users identify thin spots or over-application without pausing for individual readings.

Display, Bluetooth, Storage, and QC Mode

The user interface is centered around two 128 × 48 dot matrix LCD screens, providing visibility even when the instrument is positioned in tight or inverted orientations. For advanced documentation, the built-in Bluetooth module synchronizes with a mobile app, facilitating the creation of professional measurement reports. Operators can toggle between a “Standard Measurement Mode” for individual checks and a “QC Inspection Mode” for high-volume analysis. The onboard memory allows for the storage of 9 data sets, ensuring that critical measurements are preserved for review if a mobile device is not immediately available.

Measurement System

FeatureSpecification
ModelCoating Thickness Gauge – QualiCTG™ V
Measuring PrincipleFe: Magnetic Induction / NFe: Eddy Current
Probe TypeIntegrated with Cemented Carbide Tip
Measuring Range0.0 to 2000 μm
Measuring Unitsμm / mil
Measuring Interval0.3 seconds
Substrate DetectionAutomatic mode switching

Resolution and Accuracy

Measurement RangeResolutionAccuracy
0 to 99.9 μm0.1 μm≤ ±(2%H + 1 μm)
100 to 1000 μm1 μm≤ ±(2%H + 1 μm)
1000 to 2000 μm1 μm≤ ±3%H
Note: H is the standard value

Substrate and Environmental Limits

FeatureSpecification
Min. Measuring AreaØ = 15 mm
Min. Substrate (Fe)0.2 mm
Min. Substrate (NFe)0.05 mm
Operating Temperature0°C to 50°C
Storage Temperature-20°C to 60°C

Display, Power, and Physical

FeatureSpecification
DisplayDual-screen 128 × 48 dot matrix LCD
ConnectivityBluetooth (Mobile App Support)
Onboard Storage9 complete measurement sets
Power Supply2x 1.5V AA Alkaline batteries (DC 3V)
Dimensions90 × 65 × 28 mm
Weight123 g (with battery)

Packaging Contents

  • Coating Thickness Gauge QualiCTG™ V (1 set)
  • Fe Zero-adjustment Plate (1 pcs)
  • NFe Zero-adjustment Plate (1 pcs)
  • Standard Calibration Films (5 pcs)
  • User Manual (1 pcs)

Experience the Qualitest advantage and see how we can elevate your testing and quality control processes.

Please contact us for pricing, check availability, or ask for additional information about:

Coating Thickness Gauge – QualiCTG™ V